• J. Appl. Physiol. · Jan 1992

    Comparative Study

    Factors affecting the accuracy of esophageal balloon measurement of pleural pressure in dogs.

    • G Dechman, J Sato, and J H Bates.
    • Meakins-Christie Laboratories, McGill University, Montreal, Quebec, Canada.
    • J. Appl. Physiol. 1992 Jan 1; 72 (1): 383-8.

    AbstractSimultaneous measurement of esophageal and tracheal pressures during an occluded inspiratory effort was used to assess the accuracy of the esophageal balloon for measuring pleural pressure in dogs. Esophageal balloons were inserted in five mongrel dogs, and an occlusion test was performed with the balloon tip 5, 10, 15, 20, and 25 cm above the esophageal sphincter; at lung volumes of functional residual capacity (FRC) and FRC + 600 ml; and in supine and right- and left-side lying postures. The protocol was repeated in paralyzed animals. This time the occlusion test was performed by injecting air into a plethysmograph to change the body surface pressure, simulating pressure changes produced by respiratory efforts in spontaneously breathing animals. In 47% of the tests in spontaneously breathing dogs, the slope of esophageal vs. tracheal pressure varied greater than 10% from unity. After paralysis the slope did not vary greater than 5% from unity under any circumstance. These data indicate that the poorer performance of the occlusion test in nonparalyzed dogs is due to active tension in the walls of the esophagus and stress induced in the intrathoracic soft tissues by the descent of the diaphragm during a breathing effort.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.