    Cardiac output derived from left ventricular pressure during conductance catheter evaluations: an extended Modelflow method.

    • Sergio Valsecchi, Giovanni B Perego, Jan J Schreuder, Federica Censi, and Jos R C Jansen.
    • Medtronic Italia, Via Lucrezio Caro, Rome, Italy. sergio.valsecchi@medtronic.com
    • J Clin Monit Comput. 2007 Aug 1; 21 (4): 227-35.

    ObjectiveThe Modelflow method computes cardiac output (CO) from arterial pressure (CO-MFao) by simulating a non-linear three-element Windkessel model of aortic input impedance. We present a novel technique to apply the Modelflow method to the left ventricular pressure (Plv) signal, to obtain an estimation of CO (CO-MFlv).MethodsWe extended the model by simulating the aortic valve as a resistance placed in series to the characteristic impedance. In our model the valve resistance is infinite in diastole, while it has a patient related value during systole. Twenty one patients with heart failure were studied with a combined pressure-conductance catheter positioned in the left ventricle and a micromanometer in the aorta. CO changes were induced during temporary sequential atrio-biventricular pacing with different atrio-ventricular and inter-ventricular intervals. After a single calibration we compared CO-MFlv with CO-MFao (516 measurement series) and with CO estimated by the conductance catheter volume technique (CO-cond) (267 series).ResultsCO ranged from 2.42 to 7.59 l/min with an overall mean of 4.36 (1.38) l/min. Comparing the two Modelflow methods, the bias was -0.04 (0.36) l/min, with limits of agreement of -0.77 and 0.70 l/min, and a coefficient of variation of 8.4%. Of the 516 changes in CO from baseline values, 112 were greater than 0.5 l/min and 110 of these (98%) were scored in the same direction by both methods. For comparisons between CO-MFlv and CO-cond the bias was -0.10 (0.49), with limits of agreement of -1.12 and 0.90 l/min, and a coefficient of variation of 12.5%.ConclusionsCardiac output estimates by the modelflow method from aortic pressure and left ventricular pressure are interchangeable in patients without mitral and aortic abnormalities. After an initial calibration, CO-MFlv presents near zero bias and an adequate precision.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…