AbstractThe similarity between and the number of limb lead cables play an important role in the frequency of incorrect connection of limb electrodes. Hence, a modified electrocardiogram (ECG) acquisition procedure is proposed in this brief communication, whereby the left-leg (LL) and right-leg (RL) electrode cables are combined into 1 cable, referred to as combined LL/RL cable. The electrode wires in the combined LL/RL cable are connected to 2 electrodes placed on both sides of the LL. The combined LL/RL cable is unique enough (being thicker) not to be mistaken with the upper limb electrode cables. The proposed modification will not in any way influence the ECG waveforms or amplitudes, and it can be expected to substantially reduce incorrect limb electrode placements.