• Comput. Aided Surg. · Nov 2007

    Surgical targeting accuracy analysis of six methods for subthalamic nucleus deep brain stimulation.

    • Ting Guo, Andrew G Parrent, and Terry M Peters.
    • Imaging Research Laboratories, Robarts Research Institute, London, Ontario, Canada N6A 5K8. tguo@imaging.robarts.ca
    • Comput. Aided Surg. 2007 Nov 1; 12 (6): 325-34.

    AbstractA commonly adopted surgical target in deep brain stimulation (DBS) procedures, the subthalamic nucleus (STN) is located deep within the brain and is surrounded by delicate deep-brain structures. Symptoms of Parkinson's disease can be reduced by precisely implanting a multi-electrode stimulator at a specific location within the STN and delivering the appropriate signal to the target. A number of techniques have recently been proposed to facilitate STN DBS surgical targeting and thereby improve the surgical outcome. This paper presents a retrospective study evaluating the target localization accuracy and precision of six approaches in 55 STN DBS procedures. The targeting procedures were performed using a neurosurgical visualization and navigation system, which integrates normalized and standardized anatomical and functional information into the planning environment. In this study, we employed as the "gold standard" the actual surgical target locations determined by an experienced neurosurgeon using both pre-operative image-guided surgical target/trajectory planning and intra-operative electrophysiological exploration and confirmation. The surgical target locations determined using each of the six targeting methods were compared with the "gold standards". The average displacement between the actual surgical targets and those planned with targeting approaches was 3.0 +/- 1.3 mm, 3.0 +/- 1.3 mm, 3.0 +/- 1.0 mm, 2.6 +/- 1.1 mm, 2.5 +/- 0.9 mm, and 1.7 +/- 0.7 mm for approaches based on T2-weighted MRI, a brain atlas, T1 and T2 maps, an electrophysiological database, a collection of final surgical targets from previous patients, and the combination of these functional and anatomical data, respectively. The technique incorporating both anatomical and functional data provides the most reliable and accurate target position for STN DBS.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.