    Method for high accuracy differential pressure measurements using fluid-filled catheters.

    • Oren M Rotman, Uri Zaretsky, Avraham Shitzer, and Shmuel Einav.
    • Department of Biomedical Engineering, Faculty of Engineering, Tel Aviv University, 69978, Tel Aviv, Israel, orenrotman1@gmail.com.
    • Ann Biomed Eng. 2014 Aug 1; 42 (8): 1705-16.

    AbstractThe advantage of measuring differential pressure using fluid-filled catheters is that the system is relatively inexpensive, but the readings are not accurate and affected by the common mode pressure (CMP) distortion. High accuracy differential pressure measurements are required in various biomedical applications, such as in fluid-dynamic test rigs, or in the cath-lab, from cardiac valves efficacy to functional assessment of arterial stenoses. We have designed and built a unique system in which the pressure difference was measured along the fluid flow inside a rigid circular tube using a fluid-filled double-lumen catheter. The differential pressure measurements were taken across two side-holes near the catheter distal tip, spaced apart by 3 cm. The goal was to overcome the CMP error, which significantly distorted the output differential pressure signal and to formulate a restoration factor. A restoration formula was developed based on simultaneous gauge pressure measurements, and was tested in several different cases. Several representative cases are presented and show that the common mode artifact was reduced by factors of 12-27. The restored pressure gradient signal was validated using direct pressure drop measurements, and showed very good agreement.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…