Undersea Hyperbar M · Jan 2016
Case ReportsInaccurate pulse CO-oximetry of carboxyhemoglobin due to digital clubbing: case report.
- Nicole Harlan, Lindell K Weaver, and Kayla Deru.
- Undersea Hyperbar M. 2016 Jan 1; 43 (1): 59-61.
AbstractNewer pulse CO-oximeters provide a non-invasive and quick means of measuring oxyhemoglobin, carboxyhemoglobin and methemoglobin. Clubbing has been reported to cause inaccuracy in pulse oximeters. We present a case of inaccurate carboxy-hemoglobin measurement by pulse CO-oximetry due to digital clubbing. An 18-year-old man with a history of cystic fibrosis presented after a suicide attempt by inhalation of exhaust. At the initial emergency department evaluation, his blood carboxyhemoglobin was 33%. He was intubated, placed on 100% oxygen and transferred to our facility. Upon arrival, we placed three different pulse CO-oximeters on different fingers and toes. Carboxyhemoglobin levels measured by these meters ranged from 9%-11%. A venous blood gas drawn on arrival showed a carboxyhemoglobin level of 2.3% after four hours on 100% oxygen by endotracheal tube. Thirty minutes later, we checked arterial blood gas, which revealed a COHb level of 0.9%. Again, non-invasive carboxyhemoglobin measurements read 10%. The patient was treated with hyperbaric oxygen for carbon monoxide poisoning. This case suggests that non-invasive measurements of carboxyhemoglobin should be correlated with the clinic history and with an arterial or venous blood gas oximetry analysis.
Notes
Knowledge, pearl, summary or comment to share?