• Critical care medicine · Mar 2006

    Randomized Controlled Trial

    Double-heater-wire circuits and heat-and-moisture exchangers and the risk of ventilator-associated pneumonia.

    • Robert J Boots, Narelle George, Joan L Faoagali, John Druery, Kevin Dean, and Richard F Heller.
    • Department of Intensive Care Medicine, Royal Brisbane and Women's Hospital, Brisbane, Australia, and Department of Public Health, Manchester University, United Kingdom.
    • Crit. Care Med. 2006 Mar 1; 34 (3): 687-93.

    ObjectiveTo compare the incidence of ventilator-associated pneumonia (VAP) in patients ventilated in intensive care by means of circuits humidified with a hygroscopic heat-and-moisture exchanger with a bacterial viral filter (HME) or hot-water humidification with a heater wire in both inspiratory and expiratory circuit limbs (DHW) or the inspiratory limb only (SHW).DesignA prospective, randomized trial.SettingA metropolitan teaching hospital's general intensive care unit.PatientsThree hundred eighty-one patients requiring a minimum period of mechanical ventilation of 48 hrs.InterventionsPatients were randomized to humidification with use of an HME (n=190), SHW (n=94), or DHW (n=97).Measurements And Main ResultsStudy end points were VAP diagnosed on the basis of Clinical Pulmonary Infection Score (CPIS) (), HME resistance after 24 hrs of use, endotracheal tube resistance, and HME use per patient. VAP occurred with similar frequency in all groups (13%, HME; 14%, DHW; 10%, SHW; p=0.61) and was predicted only by current smoking (adjusted odds ratio [AOR], 2.1; 95% confidence interval [CI], 1.1-3.9; p=.03) and ventilation days (AOR, 1.05; 95% CI, 1.0-1.2; p=.001); VAP was less likely for patients with an admission diagnosis of pneumonia (AOR, 0.40; 95% CI, 0.4-0.2; p=.04). HME resistance after 24 hrs of use measured at a gas flow of 50 L/min was 0.9 cm H2O (0.4-2.9). Endotracheal tube resistance was similar for all three groups (16-19 cm H2O min/L; p=.2), as were suction frequency, secretion thickness, and blood on suctioning (p=.32, p=.06, and p=.34, respectively). The HME use per patient per day was 1.13.ConclusionsHumidification technique does not influence either VAP incidence or secretion characteristics, but HMEs may have air-flow resistance higher than manufacturer specifications after 24 hrs of use.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.