Pediatric pulmonology · Sep 1987
Performance of respirators at fast rates commonly used in neonatal intensive care units.
- A Greenough and F Greenall.
- Department of Child Health, King's College Hospital, Denmark Hill, London, England.
- Pediatr. Pulmonol. 1987 Sep 1; 3 (5): 357-61.
AbstractThe effect on tidal volume and airway pressure of increasing ventilator rate (30, 60, and 120/min) was tested in six commonly used neonatal ventilators. In all six ventilators increased flow was necessary to maintain mean airway pressure at the higher rates. Tidal volume decreased at rates of both 60 and 120/min in all six ventilators, associated with a change in pressure waveform. The most marked reduction in tidal volume, however, was associated with increased positive end-expiratory pressure (PEEP). This was only demonstrated in four ventilators, all incorporating nonassisted expiratory valves. These results stress the necessity for appropriately designed ventilators if fast rates are to be used routinely in clinical practice.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.