• Anaesth Intensive Care · Aug 1993

    Use of a new syringe pump (Springfusor) for muscle relaxant infusion.

    • C C Eagle and D F Capes.
    • Surgicentre, South Perth, Western Australia.
    • Anaesth Intensive Care. 1993 Aug 1; 21 (4): 444-6.

    AbstractA new spring-driven syringe pump (Springfusor 10, Go Medical Industries, Subiaco, W.A.) was evaluated for infusion of muscle relaxants in 50 surgical cases. After an initial bolus dose, atracurium was given at a rate of 25 to 37.5 mg/hour and vecuronium at a rate of 4 to 6 mg/hour. The mean infusion rate was found to be 4% above the nominal rate of 5 ml/hour and the accuracy in all cases was within +/- 20%. The Springfusor was found to be rugged, cheap and easy to use. It lacks dose flexibility but is suitable for continuous background infusion of muscle relaxants during surgery.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...