• Can Anaesth Soc J · Jul 1980

    Limits of high frequency percutaneous transtracheal jet ventilation using a fluidic logic controlled ventilator.

    • R B Smith, M Klain, and M Babinski.
    • Can Anaesth Soc J. 1980 Jul 1; 27 (4): 351-6.

    AbstractA study was undertaken on dogs to find the limit of carbon dioxide exchange with high frequency jet ventilation using a fluidic logic controlled oxygen jet ventilator. Fifteen dogs were ventilated through a transtracheal catheter at respiratory rates up to 600 per minute. The following were recorded: aortic, pulmonary artery, pulmonary arterial wedge, and central venous blood pressures; intratracheal pressure, electrocardiogram; inspiratory and expiratory time of the jet; arterial and central venous blood gases; intermittent cardiac output. Normal gas exchange was found up to a respiratory rate of 400 per minute with low tidal volume and low intratracheal pressures. There were no adverse circulatory effects up to a rate of 400 per minute. At rates of 500 and 600 per minute, cardiac contractility was unaffected, but a decreased heart rate and increased peripheral resistance produced a fall in cardiac output. There was no interference with the resumption of spontaneous ventilation during weaning. In a control series of five dogs, apnoeic oxygenation was used. The PaCO2 was allowed to reach 15.96 kPa (120 torr). High frequency jet ventilation was then started at a rate of 600 per minute and decreased in increments to 100 per minute. Arterial blood gases were continuously recorded through an intra-arterial catheter connected to a mass spectrometer. The PaCO2 gradually declined to normal levels as the rate decreased.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…