Arch. Dis. Child. Fetal Neonatal Ed. · Jul 2014
Comparative Study Observational StudyThe precision and accuracy of Nellcor and Masimo oximeters at low oxygen saturations (70%) in newborn lambs.
- J A Dawson, P Bastrenta, F Cavigioli, M Thio, T Ong, M L Siew, S B Hooper, and P G Davis.
- Arch. Dis. Child. Fetal Neonatal Ed. 2014 Jul 1; 99 (4): F278-81.
BackgroundAdministration of oxygen in the delivery room is informed by oxygen saturation (SpO₂). An oxygen saturation range of 60%-70% is the threshold for administering oxygen in the first minutes after birth. The accuracy of newer generation oximeters to measure SpO₂ has not been compared against the 'gold standard', direct arterial blood oxygen saturation (SaO₂) when SaO₂ is low. The aim of this study was to determine the accuracy and precision of Nellcor and Masimo oximeters to measure SpO₂ when SaO₂ <70%.MethodProspective observational study in ventilated anaesthetised newborn lambs with an indwelling carotid artery catheter. Ventilation was adjusted to achieve hypoxaemia. Nellcor (Oxi-Max 600 with Max-N sensor) and Masimo (Rad 4 with low noise optical probe (LNOP) sensor) sensors were applied to the right forelimb (preductal). An arterial blood sample was collected at 1-5 min intervals when the animal was hypoxic. The displayed SpO₂ was recorded. We used Bland-Altman analysis to determine precision and accuracy of each oximeter when SaO₂ <70%.Results17 lambs were studied, 165 measurements were obtained, 123 were SaO₂ <70%. The mean difference (±1.96 SD) Nellcor SpO₂-SaO₂ when SaO₂ <70% was 17% (-12% to 46%). The mean difference (±1.96 SD) Masimo SpO₂-SaO₂ when SaO2 <70% was 13% (-19% to 45%).ConclusionsAt SaO₂<70%, both monitors overestimated oxygen saturation (SpO₂) compared with the gold standard. Both oximeters were equally inaccurate when SaO₂ was low.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.