Comparative StudyPerformance of neonatal ventilators: the effects of changes in resistance and compliance.
- G Simbruner and G A Gregory.
- Crit. Care Med. 1981 Jul 1; 9 (7): 509-14.
AbstractThe authors tested the ability of 6 neonatal ventilators (Baby bird, Bourns BP-200, Veriflo, Bourns LS, ServoVent, SI-75) to deliver a preset tidal volume with minimal pressures when the compliance and resistance of test lung were altered. The authors also studied the effects of gas leaks from the system and the effects of PEEP. When the ventilator cycling rates exceeded 28-43/min, gas trapping occurred despite compliances and resistance appropriate for a healthy term infant. With a normal compliance and an elevated resistance, gas trapping occurred above rates of 4-6/min. Under these same conditions (normal compliance, elevated resistance), only 24-46% of the preset tidal volume was delivered. The different ventilators caused the pressure within the test lung to exceed 5, 10, and 15 cm H2O for differing amounts of time (sec)/min.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.