Case Reports
Coagulopathic-induced membrane dysfunction during extracorporeal membrane oxygenation: a case report.
- A H Stammers, L W Fristoe, K Christensen, J Deptula, R T Sydzyik, D Zavadil, and L Willett.
- Division of Clinical Perfusion Education, University of Nebraska Medical Center, Omaha 68198-5155, USA.
- Perfusion. 1997 Mar 1; 12 (2): 143-9.
AbstractThis paper describes an unusual complication of membrane dysfunction during extracorporeal membrane oxygenation (ECMO) for treatment of neonatal respiratory distress. A 2.8-kg term infant presented to our facility in severe respiratory distress and was diagnosed with primary pulmonary hypertension. After routine priming of the extracorporeal circuit, the patient was placed on veno-arterial ECMO with 8 F arterial and 12 F venous cannulae. Transfusion criteria were established which included trigger values of the following: platelet count 100,000/microliters, fibrinogen 150 mg/dl, haematocrit 40%. The ECMO course was uneventful until approximately the 132nd hour on support when the patient developed a consumptive coagulopathy, as evidenced by 55-60% reductions in both platelet count and fibrinogen concentrations, despite transfusion therapy. Total autogeneic blood product transfusion during the first 120 h of ECMO averaged 4.4 +/- 2.2 ml/h, while the transfusion rate for the final 35 h was 7.8 +/- 3.5 ml/h. Coinciding with this rise in transfusion requirements was an increase in transmembrane pressure from 0.29 to 1.52 mmHg/ml blood flow. The patient was separated from ECMO after 175 h due to a continuing coagulopathy and haemothorax. The patient was then treated with nitric oxide therapy before succumbing on the twelfth postoperative day due to refractory respiratory failure. The circuit was dissected and significant clots found in both the venous bladder and oxygenator. In addition, approximately one-third of the membrane compartment had a 'fused' circumferential pattern of dessicated clot which interrupted blood path continuity. In conclusion, this report describes an unusual complication of the ECMO oxygenator that occurred during long-term extracorporeal life support which most likely resulted from a coagulopathy.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.