Anesthesia and analgesia · May 2004
Case ReportsTracheal ring fracture during a PercuTwist tracheostomy procedure.
- Emmanuel Scherrer, Loic Tual, and Gilles Dhonneur.
- Department of Anesthesiology and Intensive Care Medicine, Henri Mondor University Hospital and Paris XII School of Medicine, Créteil, France.
- Anesth. Analg. 2004 May 1; 98 (5): 1451-3, table of contents.
UnlabelledBecause of difficult weaning from mechanical ventilation, a 59-yr-old man admitted for emergency cardiac surgery underwent a bedside PercuTwist tracheostomy on day 14 of his intensive care unit stay. We observed a double fracture of the second tracheal ring during the initial dilation process with the PercuTwist dilator, associated with distal migration of a cartilage fragment, which was avulsed from the anterior portion of the second cartilaginous ring.ImplicationsLike other antegrade single-step techniques, the PercuTwist tracheostomy presents the risk of anterior tracheal wall damage during the initial stage of the dilation process. Antegrade forces applied to the trachea should be minimized by sufficiently deep skin incision and both slow and smooth initial rotation of the dilator.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.