- O Maunz, S Penn, and A Simon.
- Department of Cardiothoracic Transplantation & Mechanical Support, Royal Brompton & Harefield NHS Trust, Harefield, UK. olaf.maunz@me.com
- Perfusion. 2013 Mar 1; 28 (2): 167-9.
AbstractWe present a case with an unusually high transmembrane pressure gradient over the oxygenator during initiation of extracorporeal membrane oxygenation (ECMO). This necessitated an immediate emergency oxygenator exchange. A detailed inspection revealed a large lipid formation in the oxygenator. We hypothesize that this lipid was dislodged during the cannulation process.