• An Pediatr (Barc) · Apr 2008

    Clinical Trial

    [The effect of helium on ventilator parameters: in vitro study with a Servoi ventilator].

    • J López-Herce Cid, J Urbano Villaescusa, E Cidoncha Escobar, J Del Castillo Peral, M J Santiago Lozano, S Mencía Bartolomé, and J M Bellón Cano.
    • Sección de Cuidados Intensivos Pediátricos. Hospital General Universitario Gregorio Marañón. Madrid. España. pielvi@ya.com
    • An Pediatr (Barc). 2008 Apr 1; 68 (4): 336-41.

    ObjectiveTo analyze the changes in respiratory parameters in a Servoi ventilator with heliox.Material And MethodsIn vitro study with a Servoi (Maquet) ventilator in volume controlled (VC), pressure controlled (PC) and volume control regulated by pressure (VCPR) modes connected to an artificial lung. A heliox tank with a fixed concentration of helium 70 % and oxygen 30 % was connected to the air inlet of the ventilator. The ventilator was set in VC mode with tidal volumes of 30, 50, 100, 250 and 500 ml; in PC mode with pressure of 20 and 30 cmH(2)O and in VCPR mode with tidal volume of 150 ml. In each case FiO(2) of 21, 30, 40, 50, 60, 70, 80, 90 and 100 % was used. The FiO(2), inspired and expired tidal volume and inspiratory pressure measured by the ventilator and a pitot spirometer (Datex_S5) were compared.ResultsIn VC and VCPR modes the increase in helium produced a progressive decrease in the pressure needed to administrate the set volume. Heliox also produced a decrease in the tidal volume measured by the spirometer and the tidal expired volume measured by the respirator. In PC mode, heliox produced a progressive increase in the inspired tidal volume, increasing the differences between inspired and expired tidal volumes.ConclusionsHeliox used with Servoi ventilator produces a decrease in inspiratory pressures in VC and VCPR modes, and an increase in inspiratory tidal volume in PC mode. In all modes heliox reduced the expired tidal volume measured by the ventilator and the spirometer. These changes should be borne in mind if heliox is used with this ventilator.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.