Anesthesia and analgesia · Aug 2010
Comparative StudyNegative intratracheal pressure produced by esophageal detector devices causes tracheal wall collapse in a porcine cardiac arrest model.
- Jason R Schuh, Robert Bettendorf, David Shapiro, David L Trowbridge, Eugenio Lujan, Patrick Boyle, Sean Birmingham, Rhodel Dacanay, and Richard Green.
- Department of Anesthesiology and Post-Anesthesia Care, Naval Medical Center, San Diego, CA 92134-1005, USA.
- Anesth. Analg. 2010 Aug 1; 111 (2): 468-72.
BackgroundEsophageal detector devices (EDDs) impose negative pressure on the trachea or esophagus to verify endotracheal tube (ETT) position. In cardiac arrest, the smooth muscle of the lower esophageal sphincter relaxes in a time-dependent and irreversible manner. If relaxation also occurs in the muscular posterior tracheal wall, it could predispose tracheal invagination or collapse with negative pressure, potentially yielding false-negative (tracheal ETT, EDD indicates esophagus) results. We compared 3 different EDDs in their ability to correctly discriminate tracheal and esophageal intubation.MethodsETTs were placed into the trachea and esophagus of 5 domestic swine, and bronchoscopy was used to visualize the trachea while 3 EDDs were tested. Tracheal wall activity was observed before and after induced cardiac arrest. Tracheal ETTs were aspirated with increasing negative force and pressures at initial wall movement and >50% tracheal lumen obliteration were recorded. Measurements were repeated at 4, 8, and 12 minutes postarrest and pressures at tracheal wall collapse pre- and postarrest were determined. EDDs were also tested on esophageal ETTs prearrest and at 6 and 10 minutes postarrest.ResultsIn a closed system, each EDD generated more than -100 cm H(2)O pressure. Average prearrest pressure at tracheal collapse was -112 cm H(2)O. Average postarrest collapse pressures were -68, -66, and -54 cm H(2)O at 4, 8, and 12 minutes postarrest. One EDD consistently gave equivocal results; the remaining 2 gave accurate results in all subjects. Most observed movement was insufficient to cause device failure although tracheal wall movement was noted in all postarrest EDD trials. Esophageal intubation was correctly determined at all times pre- and postarrest.ConclusionThese findings describe a mechanism for false-negative results from decreased posterior tracheal wall tone during cardiac arrest. Further studies are required to elucidate factors contributing to its occurrence and impact on EDD use.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.