    Evaluating the efficiency of desflurane reflection in two commercially available reflectors.

    • Hagen Bomberg, Marcel Wessendorf, Martin Bellgardt, Max Veddeler, Stefan Wagenpfeil, Thomas Volk, Heinrich V Groesdonk, and Andreas Meiser.
    • Department of Anesthesiology, Intensive Care Medicine and Pain Medicine Saarland University Medical Center, University of Saarland, Kirrbergerstrasse 1, 66421, Homburg/Saar, Germany. hagen.bomberg@uniklinikum-saarland.de.
    • J Clin Monit Comput. 2018 Aug 1; 32 (4): 605-614.

    AbstractWith the AnaConDa™ and the MIRUS™ system, volatile anesthetics can be administered for inhalation sedation in intensive care units. Instead of a circle system, both devices use anesthetic reflectors to save on the anesthetic agent. We studied the efficiency of desflurane reflection with both devices using different tidal volumes (VT), respiratory rates (RR), and 'patient' concentrations (CPat) in a bench study. A test lung was ventilated with four settings (volume control, RR × VT: 10 × 300 mL, 10 × 500 mL, 20 × 500 mL, 10 × 1000 mL). Two different methods for determination of reflection efficiency were established: First (steady state), a bypass flow carried desflurane into the test lung (flowin), the input concentration (Cin) was varied (1-17 vol%), and the same flow (flowex, Cex) was suctioned from the test lung. After equilibration, CPat was stored online and averaged; efficiency [%] was calculated [Formula: see text]. Second (washout), flowin and flowex were stopped, the decline of CPat was measured; efficiency was calculated from the decay constant of the exponential regression equation. Both measurement methods yielded similar results (Bland-Altman: bias: -0.9 %, accuracy: ±5.55 %). Efficiencies higher than 80 % (>80 % of molecules exhaled are reflected) could be demonstrated in the clinical range of CPat and VT. Efficiency inversely correlates with the product of CPat and VT which can be imagined as the volume of anesthetic vapor exhaled by the patient in one breath, but not with the respiratory frequency. Efficiency of the AnaConDa™ was higher for each setting compared with the MIRUS™. Desflurane is reflected by both reflectors with efficiencies high enough for clinical use.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…