Randomized Controlled Trial Comparative Study
Prospective randomized study comparing the clinical outcomes between inferior vena cava Greenfield and TrapEase filters.
- Fred Usoh, Anil Hingorani, Enrico Ascher, Alexander Shiferson, Nirav Patel, Kapil Gopal, Natalia Marks, and Theresa Jacob.
- Division of Vascular Services, Maimonides Medical Center, Brooklyn, NY 11219, USA.
- J. Vasc. Surg. 2010 Aug 1; 52 (2): 394-9.
ObjectiveAlthough anticoagulation remains the mainstay of treatment for deep venous thrombosis, the use of inferior vena cava (IVC) filters when anticoagulation has failed or when contraindicated remains a safe and effective treatment. Greenfield (Boston Scientific, Natick, Mass) and TrapEase (Cordis, Bridgewater, NJ) filters are arguably among the most popular filtration devices. The Greenfield filter (12F introducer) has been in use for >30 years and has been well studied. The TrapEase filter (6F introducer) has been used since 2000, with a limited number of studies. Good guidelines to help determine which filter to use in any given situation are lacking; therefore, this randomized study prospectively compared the clinical outcomes (access-site thrombosis, filter thrombosis, and symptomatic pulmonary embolism [PE]) between these filters.MethodsBetween July 2006 and November 2008, 156 patients (63 men, 93 women; mean age, 75 years; range, 38-101 years) were randomized: 84 to Greenfield and 72 to TrapEase IVC filter insertion in the infrarenal position using angiographic guidance. Postoperative follow-up comprised serial lower extremity and IVC/iliac vein (IV) duplex imaging (78.2%) at day 1, week 1, every 3 months for the first year, and every 6 months for the second year; clinical evaluation, and clinic visits. During this period, 349 patients (143 men, 206 women; mean age, 75 years; range, 24-96 years) were not randomized.ResultThe indications for filter placement, in the 156 randomized patients, were gastrointestinal bleeding, 37; intracranial hemorrhage, 12; free-floating clot, 19; failure of anticoagulation, 29; PE, 27; prophylactic, 4; and others, 32. During a mean 12-month follow-up (range, 0-39 months), symptomatic IVC/IV thrombosis developed in five patients (6.94%) in the TrapEase group and none in the Greenfield group (P = .019). No filter migration, access-site thrombosis, misplacement, or IVC perforation occurred. Recurrent PE was suspected in one of the five patients with IVC/IV thrombosis. Overall mortality was 42.3% (66 patients), and 30-day mortality was 13.5% (21 patients: 10 TrapEase, 11 Greenfield). The study was initially designed to recruit 360 patients in both TrapEase and Greenfield filters in 2 years to demonstrate any statistical significance but was prematurely concluded due to the interim results.ConclusionA higher rate of symptomatic IVC/IV thrombosis is associated with TrapEase filter placement. However, the TrapEase filter still has a selective clinical role in the prevention of thromboembolism in selected patients who are coagulopathic. This is the first randomized prospective study comparing IVC filters since their inception in 1967.Copyright (c) 2010 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.