Comparative Study
[Comparative study on the strength of different mechanisms of operation of multidirectionally angle-stable distal radius plates].
- S Rausch, K Hoffmeier, B G Gueorguiev, K Klos, F Gras, G O Hofmann, and T Mückley.
- Klinik für Unfall-, Hand- und Wiederherstellungschirurgie, Klinikum der Friedrich-Schiller-Universität Jena, Erlanger Allee 102, Jena. sascha.rausch@med.uni-jena.de
- Z Orthop Unfall. 2011 Dec 1; 149 (6): 694-8.
IntroductionPolyaxial angle-stable plating is thought to be particularly beneficial in the management of complex intra-articular fractures of the distal radius. The present study was performed to investigate the strength of polyaxial locking interfaces of distal radius plates.Material And MethodsWe tested the polyaxial interfaces of 3 different distal radius plates (2.4 mm Variable Angle LCP Two-Column Volar Distal Radius Plate, Synthes, Palmar Classic, Königsee Implantate and VariAx Plate Stryker). The strength of 0° and 10° screw locking angle was obtained during static loading.ResultsThe strength of Palmar Classic with a 0° locking angle is significantly the best of all tested systems. With a 10° locking angle there is no significant difference between Palmar Classic, Two column Plate and VariAx Plate.ConclusionThe strength of polyaxial interfaces differs between the tested systems. A reduction of ultimate strength is due to increases of screw locking angle. The design of polyaxial locking interfaces should be investigated in human bone models.© Georg Thieme Verlag KG Stuttgart · New York.