• Comput Methods Biomech Biomed Engin · Feb 2011

    Comparative Study

    Experimental and probabilistic analysis of distal femoral periprosthetic fracture: a comparison of locking plate and intramedullary nail fixation. Part A: experimental investigation.

    • Christina Salas, Deana Mercer, Thomas A DeCoster, and Mahmoud M Reda Taha.
    • Department of Mechanical Engineering and Department of Civil Engineering, The University of New Mexico, Albuquerque, NM 87131, USA. csalas@unm.edu
    • Comput Methods Biomech Biomed Engin. 2011 Feb 1; 14 (2): 157-64.

    AbstractThe following is a two-part study. Part A evaluates biomechanically intramedullary (IM) nails vs. locking plates for fixation of femoral fractures in osteoporotic bone. Part B of this study introduces a deterministic finite element model of each construct type and investigates the probability of periprosthetic fracture of the locking plate compared with the retrograde IM nail using Monte Carlo simulation. For Part A, an extra-articular, metaphyseal wedge fracture pattern was created in 11 osteoporotic fourth-generation composite femurs. Fixation was performed with a locking plate or a retrograde IM nail. Axial, torsion and bending cyclic loading to simulate post-operative damage accumulation were performed followed by ramped load to failure. Locking plates proved to be more stable (using stiffness as the determining factor) in osteoporotic bone as observed under low load cycle conditions. However, some of these advantages were offset by a greater incidence of sudden periprosthetic fracture observed under ramped loading conditions. Cadaveric, osteoporotic femurs included as a case study also exhibited periprosthetic fracture, but failure was accompanied by catastrophic comminution of the cortex. Periprosthetic failure at the implant end including bone comminution is difficult to salvage with revision fixation. The weakened trabecular matrix and thinned cortex of osteoporotic bone may increase the incidence of periprosthetic fracture. It is, therefore, essential for the surgeon to consider all possible loading scenarios when recommending an ideal implant for the osteoporotic patient.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.