Case Reports
Misplacement of a right internal jugular vein haemodialysis catheter into the mediastinum.
- M K H Tong, Y P Siu, Y Y Ng, T H Kwan, and T C Au.
- Division of Nephrology, Department of Medicine, Tuen Mun Hospital, Tuen Mun, Hong Kong. khmtong@netvigator.com
- Hong Kong Med J. 2004 Apr 1; 10 (2): 135-8.
AbstractA 69-year-old woman with end-stage renal failure discontinued continuous ambulatory peritoneal dialysis and commenced temporary haemodialysis because of resistant peritonitis. Right internal jugular vein haemodialysis catheter placement was performed. The cuffed, tunnelled haemodialysis catheter was inserted using the modified Seldinger technique. When haemodialysis was initiated the following day, blood could not be aspirated from the catheter and the patient complained of central chest pain during the aspiration. Subsequent venography and computed tomography scan of the thorax showed that the catheter was placed extraluminally into the posterior mediastinum. The importance of a chest radiograph after placement of a central venous catheter is highlighted by this case report. Subtle deviations in catheter position from normal should alert the physician to the possibility of catheter misplacement and lead to further investigation.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.