Randomized Controlled Trial Multicenter Study
Sensor-driven position-adaptive spinal cord stimulation for chronic pain.
- David M Schultz, Lynn Webster, Peter Kosek, Urfan Dar, Ye Tan, and Mark Sun.
- MAPS Applied Research Center, Edina, MN 55435, USA. dschultz@painphysicians.com
- Pain Physician. 2012 Jan 1;15(1):1-12.
BackgroundVariation in the intensity of neurostimulation due to body position is a practical problem for many patients implanted with spinal cord stimulation (SCS) systems because positional changes may result in overstimulation or understimulation that leads to frequent need for compensatory manual programming adjustments.ObjectivesThe purpose of this study was to assess the safety and effectiveness of a novel type of SCS therapy designed to automatically adapt stimulation amplitude in response to changes in a patient's position or activity.The primary objective of the study was to demonstrate that automatic position adaptive SCS benefited patients in terms of pain relief and/or convenience compared with neurostimulation adjusted with conventional manual programming. Secondary objectives included assessment of worsened pain relief with automatic adjustment; change in pain score; and the number of manual programming adjustments with position-adaptive neurostimulation compared with manual programming.Study DesignProspective, multicenter, open-label, randomized crossover study.SettingTen interventional pain management centers in the US.MethodsPatients were enrolled a minimum of one week after a successful SCS screening trial. They were then implanted with the RestoreSensor neurostimulation device (Medtronic, Inc., Minneapolis, MN) that could be programmed to either automatic position-adaptive stimulation (AdaptiveStim) or manual adjustment of stimulation parameters. After implant, all devices were programmed to conventional manual adjustment for a 4-week postoperative period. The patients were then randomized to either conventional manual programming adjustment or position-adaptive stimulation with crossover to the opposite treatment arm occurring at 6 weeks after randomization. The patients were followed for another 6 weeks after crossover. This study was conducted under an FDA-approved Investigational Device Exemption (IDE) and approval of the responsible Institutional Review Boards (IRBs) of the study centers.ResultsSeventy-nine patients were enrolled in the study. In an intent-to-treat analysis, 86.5% of patients achieved the primary objective of improved pain relief with no loss of convenience or improved convenience with no loss of pain relief using automatic position-adaptive stimulation compared with using conventional manual programming adjustment alone. This was statistically significantly greater than the predefined minimum success rate of 25%, p < 0.001 (exact one-sided 97.5% lower confidence limit was 76.5%). Only 2.8% of patients reported worsened pain relief during position-adaptive stimulation compared with manual programming. There was a statistically significant reduction in the mean numeric pain rating scale score compared with baseline scores in both treatment arms. Additionally, position-adaptive stimulation demonstrated a statistically significant 41% reduction in the daily average number of programming button presses for amplitude adjustment compared with manual programming (18.2 per day versus 30.7 per day, P = 0.002). Functional improvements reported with position-adaptive stimulation included: improved comfort during position changes (80.3%); improved activity (69%); and improved sleep (47.9%). Adverse events associated with uncomfortable sensations from stimulation did not differ significantly between treatment arms. The incidence of device-related serious adverse events was 3.9%.LimitationsPatients and physicians were not blinded to whether devices were programmed to automatic position-adaptive stimulation or manual adjustment. Responses to assessment questionnaires were based on patient recall.ConclusionsThe study demonstrated that automatic position-adaptive stimulation is safe and effective in providing benefits in terms of patient-reported improved pain relief and convenience compared with using manual programming adjustment alone.Clinical TrialNCT01106404.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.