J Clin Monit Comput · Oct 2017
Evaluation of a non-invasive multisensor accelerometer for calculating energy expenditure in ventilated intensive care patients compared to indirect calorimetry and predictive equations.
- Janine Krüger, Matthias Kraft, Matthias Gründling, Sigrun Friesecke, Simone Gärtner, Lena J Vogt, Nicole Schüler, Antje Steveling, Markus M Lerch, and Ali A Aghdassi.
- Department of Medicine A, University Medicine Greifswald, Ferdinand-Sauerbruch-Straße, 17475, Greifswald, Germany.
- J Clin Monit Comput. 2017 Oct 1; 31 (5): 1009-1017.
AbstractContinuous measurement of resting energy expenditure (REE) in critically ill patients remains challenging but is required to prevent malnutrition. SenseWear Pro 3 Armband (SWA) is a research grade accelerometer for assessment of REE with the advantage of easy handling. In a prospective study we compared SWA with indirect calorimetry (IC) and predictive equations in critically ill, ventilated patients. REE was measured by SWA, IC and calculated by predictive formulas. Potential confounding factors that influence REE were also recorded. Results of SenseWear Armband and indirect calorimetry were compared using the Bland-Altman method. 34 ICU patients were investigated. SWA underestimated resting energy expenditure compared to IC with a mean bias of ΔREE = -253.6 ± 333.2 kcal, equivalent to -11.7 % (p = 0.025). This underestimation was seen in both, medical (-14.9 %) and surgical (-12.9 %) patients and the bias was greater in patients with fever (-19.0 %), tachycardia (-18.7 %) or tachypnea (-26.2 %). Differences were also noted when SWA was compared to predictive formulas. At present, SWA cannot be regarded as an alternative to indirect calorimetry. Individual measurements are often inaccurate and should be used with caution until improved algorithms, based on the results of this study, have been implemented.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.