-
Anesthesia and analgesia · Oct 2016
Clinical Trial Observational StudyThe Feasibility of a Completely Automated Total IV Anesthesia Drug Delivery System for Cardiac Surgery.
- Cedrick Zaouter, Thomas M Hemmerling, Romain Lanchon, Emanuela Valoti, Alain Remy, Sébastien Leuillet, and Alexandre Ouattara.
- From the *CHU de Bordeaux, Service d Anesthésie-Réanimation II, Bordeaux, France; †Division of Experimental Surgery, Department of Anaesthesia, Arnold and Blema Steinberg Simulation Centre, McGill University, Montreal, Canada; ‡Servizio di Anestesia Rianimazione e terapia intensiva, Università degli Studi di Pavia, Dip. Scienze Clinico-Chirurgiche Diagnostiche e Pediatriche, Unità di Anestesiologia - Fondazione IRCCS Policlinico San Matteo, Pavia, Italy; and §Biofortis, Mérieux NutriSciences Company, Saint-Herblain, France.
- Anesth. Analg. 2016 Oct 1; 123 (4): 885-93.
BackgroundIn this pilot study, we tested a novel automatic anesthesia system for closed-loop administration of IV anesthesia drugs for cardiac surgical procedures with cardiopulmonary bypass. This anesthesia drug delivery robot integrates all 3 components of general anesthesia: hypnosis, analgesia, and muscle relaxation.MethodsTwenty patients scheduled for elective cardiac surgery with cardiopulmonary bypass were enrolled. Propofol, remifentanil, and rocuronium were administered using closed-loop feedback control. The main objective was the feasibility of closed-loop anesthesia defined as successful automated cardiac anesthesia without manual override by the attending anesthesiologist. Secondary qualitative observations were clinical and controller performances. The clinical performance of hypnosis control was the efficacy to maintain a bispectral index (BIS) of 45. To evaluate the hypnosis performance, BIS values were stratified into 4 categories: "excellent," "good," "poor," and "inadequate" hypnosis control defined as BIS values within 10%, ranging from 11% to 20%, ranging from 21% to 30%, or >30% of the target value, respectively. The clinical performance of analgesia was the efficacy to maintain NociMap values close to 0. The analgesia performance was assessed classifying the NociMap values in 3 pain control groups: -33 to +33 representing excellent pain control, -34 to -66 and +34 to +66 representing good pain control, and -67 to -100 and +67 to +100 representing insufficient pain control. The controller performance was calculated using the Varvel parameters.ResultsRobotic anesthesia was successful in 16 patients, which is equivalent to 80% (97.5% confidence interval [CI], 53%-95%) of the patients undergoing cardiac surgery. Four patients were excluded from the final analysis because of technical problems with the automated anesthesia delivery system. The secondary qualitative observations revealed that the clinical performance of hypnosis allowed an excellent and good control during 70% (97.5% CI, 63%-76%) of maintenance time and an insufficient clinical performance of analgesia for only 3% (97.5% CI, 1%-6%) of maintenance time.ConclusionsThe completely automated closed-loop system tested in this investigation could be used successfully and safely for cardiac surgery necessitating cardiopulmonary bypass. The results of the present trial showed satisfactory clinical performance of anesthesia control.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.