- Laura Gochicoa-Rangel, Rogelio Pérez-Padilla, Juan Carlos Vázquez-García, Mónica Silva-Cerón, Silvia Cid-Juárez, David Martínez-Briseño, Paul L Enright, Robert Jensen, and Luis Torre-Bouscoulet.
- Departamento de Fisiología Respiratoria, Instituto Nacional de Enfermedades Respiratorias "Ismael Cosío Villegas," México, Distrito Federal, México.
- Respir Care. 2017 Feb 1; 62 (2): 231-235.
BackgroundThe 2005 American Thoracic Society/European Respiratory Society guidelines for single-breath diffusing capacity of the lung for carbon monoxide (DLCO) recommend a weekly biological control test and/or DLCO simulator to detect instrument error drift. Very little has been published regarding the results of such a quality assurance program. Our aim was to analyze the long-term stability of a portable DLCO instrument.MethodsWe used a new EasyOne Pro system and checked its accuracy using a DLCO simulator with 2 reference gases (concentration A: carbon monoxide [CO] = 0.1% and helium = 6.52%; concentration B: CO = 0.08% and helium = 7.21%) during the first 3 y of use in our large clinical laboratory. To detect instrument drift, a healthy woman (MSC), age 43 y old at baseline, tested herself every week during this period of time.ResultsMore than 6,000 spirometry and 5,000 DLCO maneuvers were done using this instrument for patients during these 3 y. There were no failures in the daily volume and flow checks or the CO and helium calibration checks performed automatically by the instrument. The differences between the simulator DLCO and the measured DLCO were -0.91 ± 1.33 mL/min/mm Hg and -0.61 ± 1.45 mL/min/mm Hg for concentration A and concentration B, respectively. The results of the 110 biological control tests were: mean 30.8 ± 1.7 mL/min/mm Hg (95% CI 30.5-31.1), coefficient of variation of 5.4% in DLCO, and repeatability of 2.5 mL/min/mm Hg. Only 4 measurements were outside ±3 mL/min/mm Hg (3.6%). Her mean alveolar volume was 4.2 ± 0.25 L with coefficient of variation of 6.2%; her inspired volume was 3.05 ± 0.14 L, and coefficient of variation = 4.5%.ConclusionsMeasurements of DLCO were stable over the 3-y period without any need for manual recalibration of the instrument. The biological control was as good as the DLCO simulator to evaluate this kind of device in a long-term laboratory quality control program.Copyright © 2017 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.