    Validation of an indirect calorimeter using n-of-1 methodology.

    • David C Frankenfield, Christine M Ashcraft, Carola Wood, and Vernon M Chinchilli.
    • Department of Clinical Nutrition, Penn State Milton S. Hershey Medical Center, 500 University Drive, Hershey, PA 17033, United States; Department of Nursing, Penn State Milton S. Hershey Medical Center, 500 University Drive, Hershey, PA 17033, United States. Electronic address: Dfrankenfield@hmc.psu.edu.
    • Clin Nutr. 2016 Feb 1; 35 (1): 163-168.

    BackgroundValidation data for currently available indirect calorimeters is limited. The purpose of this investigation was to validate a newer indirect calorimeter system (Vmax Encore) against a criterion device (Deltatrac Metabolic Monitor) in spontaneously breathing mode.Materials And MethodsAn n-of-1 methodology was used in which both indirect calorimeters were employed in the same subject repeatedly until 15 measurement pairs were generated for oxygen consumption (VO2), carbon dioxide production (VCO2), resting metabolic rate (RMR), and respiratory quotient (RQ). Bias was defined as a 95% confidence interval of differences between devices that excluded zero. A set of intradevice measurement pairs was also obtained to analyze the precision of the Vmax and Deltatrac (precision defined as not more than 5% of measurements being ≥5% different).ResultsBoth Vmax and Deltatrac precisely measured VO2 and RMR. The largest intradevice difference for RMR in the Deltatrac was 4.7% (one of 20 pairs) and in the Vmax 4.8%. On the other hand, VCO2 and RQ were not measured precisely by either device, with 10% or more of the measurement pairs differing by ≥ 5%. The Vmax was biased toward low measurement of VO2 and RMR relative to the Deltatrac. Oxygen consumption was on average 4.5 ± 2.4% lower for the Vmax device compared to Deltatrac while RMR was lower by an average of 4.1 ± 2.2%.ConclusionsThe Vmax Encore indirect calorimeter is a precise instrument for measuring VO2 and RMR in spontaneously breathing individuals, but it is biased toward lower values compared to the Deltatrac.Copyright © 2015 Elsevier Ltd and European Society for Clinical Nutrition and Metabolism. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.