- J P Bongard, C Pahud, and R De Haller.
- Centre Antituberculeux, Hôpital Cantonal Universitaire, Geneve, Switzerland.
- Eur. Respir. J. 1989 Mar 1; 2 (3): 280-2.
AbstractThe oxygen concentration (O2%) produced by 12 type A concentrators, with a working time of 28-18,099 h, and 6 type B concentrators, with a working time of 0-3,033 h, was measured over a 12 month period in the user's home, at a flow rate of 2 l.min-1. One hundred and two measurements of O2% (mean 82.9), made at least once monthly by a visiting nurse, showed that type A concentrators were usually delivering less than 92% expected O2. Four concentrators were delivering less than 40% O2 after a working time of only 4,000 h. Sixty two measurements made by the manufacturer confirmed these findings (mean 86.8). The 18 measurements performed by the nurse on the type B concentrators showed expected O2% values (mean 93.1). Our study demonstrates the necessity of regular clinical and technical surveillance, at the user's home, during long-term domiciliary oxygen therapy and the need for the manufacturer to incorporate an alarm system monitoring the O2% into oxygen concentrators.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.