• J. Clin. Microbiol. · May 2015

    Laboratory evaluation of the Liat HIV Quant (IQuum) whole-blood and plasma HIV-1 viral load assays for point-of-care testing in South Africa.

    • Lesley Scott, Natasha Gous, Sergio Carmona, and Wendy Stevens.
    • Department of Haematology and Molecular Medicine, School of Pathology, Faculty of Health Science, University of Witwatersrand, Johannesburg, South Africa Lesley.Scott@nhls.ac.za.
    • J. Clin. Microbiol. 2015 May 1; 53 (5): 1616-21.

    AbstractPoint-of-care (POC) HIV viral load (VL) testing offers the potential to reduce turnaround times for antiretroviral therapy monitoring, offer near-patient acute HIV diagnosis in adults, extend existing centralized VL services, screen women in labor, and prompt pediatrics to early treatment. The Liat HIV Quant plasma and whole-blood assays, prerelease version, were evaluated in South Africa. The precision, accuracy, linearity, and agreement of the Liat HIV Quant whole-blood and plasma assays were compared to those of reference technologies (Roche CAP CTMv2.0 and Abbott RealTime HIV-1) on an HIV verification plasma panel (n = 42) and HIV clinical specimens (n = 163). HIV Quant plasma assay showed good performance, with a 2.7% similarity coefficient of variation (CV) compared to the Abbott assay and a 1.8% similarity CV compared to the Roche test on the verification panel, and 100% specificity. HIV Quant plasma had substantial agreement (pc [concordance correlation] = 0.96) with Roche on clinical specimens and increased variability (pc = 0.73) in the range of <3.0 log copies/ml range with the HIV Quant whole-blood assay. HIV Quant plasma assay had good linearity (2.0 to 5.0 log copies/ml; R(2) = 0.99). Clinical sensitivity at a viral load of 1,000 copies/ml of the HIV Quant plasma and whole-blood assays compared to that of the Roche assay (n = 94) was 100% (confidence interval [CI], 95.3% to 100%). The specificity of HIV Quant plasma was 88.2% (CI, 63.6% to 98.5%), and that for whole blood was 41.2% (CI, 18.4% to 67.1%). No virological failure (downward misclassification) was missed. Liat HIV Quant plasma assay can be interchanged with existing VL technology in South Africa. Liat HIV Quant whole-blood assay would be advantageous for POC early infant diagnosis at birth and adult adherence monitoring and needs to be evaluated further in this clinical context. LIAT cartridges currently require cold storage, but the technology is user-friendly and robust. Clinical cost and implementation modeling is required.Copyright © 2015, American Society for Microbiology. All Rights Reserved.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…