• Journal of anesthesia · Feb 2017

    Comparative Study

    Effect of posture and extracranial contamination on results of cerebral oximetry by near-infrared spectroscopy.

    • Shinya Kato, Kenji Yoshitani, Yosuke Kubota, Yuzuru Inatomi, and Yoshihiko Ohnishi.
    • Department of Anesthesiology, National Cerebral and Cardiovascular Center, Fujishirodai, Suita, Osaka, 565-8565, Japan.
    • J Anesth. 2017 Feb 1; 31 (1): 103-110.

    BackgroundNear-infrared spectroscopy (NIRS)-based cerebral oximetry is a noninvasive technology used to estimate regional cerebral oxygen saturation (rSO2). Extracranial blood flow is known to substantially affect rSO2 values measured by most clinically available devices. Several studies have also reported that the Trendelenburg position and upright position have a larger effect on rSO2 measurements than the supine position. Therefore, we investigated the effect of these two positions (the Trendelenburg position versus the upright position) and extracranial contamination on rSO2 measurements obtained using two commercially available devices and one prototype device.MethodsTwelve healthy volunteers were enrolled in the study. They each had three cerebral oximetry devices applied to their forehead (FORE-SIGHT ELITE™, CAS Medical Systems Inc., Branford, CT, USA; INVOS 5100c™, Medtronic, Minneapolis, MN, USA; and NIRO-TRS, Hamamatsu Photonics, Hamamatsu, Japan). A circumferential pneumatic head cuff was positioned proximal to the NIRS cerebral oximetry sensors. We measured rSO2, heart rate (HR), and blood pressure (BP) in six conditions (supine, Trendelenburg and upright positions, with and without scalp ischemia induced by head cuff inflation) every 5 min with each oximetry device. Total hemoglobin (tHb), which is associated with cerebral blood volume (CBV) as measured by positron emission tomography, was measured using the NIRO-TRS device to determine extracranial blood volume in each position.ResultsMeasurements of rSO2 with all the devices were affected by extracranial contamination. The percentage of extracranial contamination was highest with INVOS 5100c™ in the upright position (INVOS, 21.3%; FORE-SIGHT, 14.3%; NIRO-TRS, 3.6%). Measurements of rSO2 obtained in the upright position were significantly lower than those obtained in the supine position, using INVOS-5100c™ and FORE-SIGHT ELITE™ (71 vs. 74% and 67 vs. 72%, respectively), but not using NIRO-TRS (62 vs. 64%). A significant decrease in tHb was observed after head cuff inflation in the supine and Trendelenburg positions (supine, 0.132-0.123 μmol/l; Trendelenburg, 0.133-0.125 μmol/l).ConclusionsExcept when using NIRO-TRS, measurements of rSO2 in the forehead are significantly lower when measured in the upright position than in the supine position. All devices in this study were affected by extracranial contamination.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.