    Comparison of pneumotachography and anemometery for flow measurement during mechanical ventilation with volatile anesthetics.

    • Jarred R Mondoñedo, Jacob Herrmann, John S McNeil, and David W Kaczka.
    • Department of Biomedical Engineering and School of Medicine, Boston University, 44 Cummington Mall, Boston, MA, 02215, USA.
    • J Clin Monit Comput. 2017 Dec 1; 31 (6): 1263-1271.

    AbstractVolatile anesthetics alter the physical properties of inhaled gases, such as density and viscosity. We hypothesized that the use of these agents during mechanical ventilation would yield systematic biases in estimates of flow ([Formula: see text]) and tidal volume (V T) for two commonly used flowmeters: the pneumotachograph (PNT), which measures a differential pressure across a calibrated resistive element, and the hot-wire anemometer (HWA), which operates based on convective heat transfer from a current-carrying wire to a flowing gas. We measured [Formula: see text] during ventilation of a spring-loaded mechanical test lung, using both the PNT and HWA placed in series at the airway opening. Delivered V T was estimated from the numerically-integrated [Formula: see text]. Measurements were acquired under baseline conditions with room air, and during ventilation with increasing concentrations of isoflurane, sevoflurane, and desflurane. We also evaluated a simple compensation technique for HWA flow, which accounted for changes in gas mixture density. We found that discrepancies in estimated V T between the PNT and HWA occurred during ventilation with isoflurane (6.3 ± 3.0%), sevoflurane (10.0 ± 7.3%), and desflurane (25.8 ± 17.2%) compared to baseline conditions. The magnitude of these discrepancies increased with anesthetic concentration. A simple compensation factor based on density reduced observed differences between the flowmeters, regardless of the anesthetic or concentration. These data indicate that the choice and concentration of anesthetic agents are primary factors for differences in estimated V T between the PNT and HWA. Such discrepancies may be compensated by accounting for alterations in gas density.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.