    Performance of the PneuX System: A Bench Study Comparison With 4 Other Endotracheal Tube Cuffs.

    • Christopher T Chenelle, Taiga Itagaki, Daniel F Fisher, Lorenzo Berra, and Robert M Kacmarek.
    • Department of Respiratory Care, Massachusetts General Hospital, Boston, Massachusetts.
    • Respir Care. 2017 Jan 1; 62 (1): 102-112.

    BackgroundCuff design affects microaspiration, a risk factor for pneumonia. We questioned whether the PneuX low-volume fold-free cuff design would prevent cuff leakage and maintain the same tracheal wall pressure as high-volume, low-pressure (HVLP) cuffs.MethodsWe evaluated 4 HVLP-cuffed endotracheal tubes (ETTs), Hi-Lo (polyvinyl chloride [PVC]), Microcuff (polyurethane [PU]), SealGuard (PU + tapered), and TaperGuard (PVC + tapered), and the PneuX with its dedicated tracheal seal monitor. In Part 1, we determined tracheal wall pressure using each cuff's capacity to support water columns across recommended intracuff pressures. In Part 2, we evaluated the tracheal seal monitor function at recommended settings. In Part 3, we compared leakage volumes of all ETTs during 30 min of simulated mechanical ventilation or during 8 h if no leak occurred. Parts 1 and 3 were performed with/without lubrication and PEEP.ResultsIn Part 1, PneuX cuffs exerted an average tracheal wall pressure of 27.4 ± 2.4 cm H2O at the recommended intracuff pressure of approximately 80 cm H2O. Tracheal wall pressure did not differ among HVLP cuffs (19.6 ± 1.4 to 29.5 ± 1.4 cm H2O). In Part 2, preinflation intracuff pressure affected the time to obtain tracheal seal monitor pressure attainment (P < .01). The tracheal seal monitor generated average calculated tracheal wall pressure of 33.4 ± 1.2 cm H2O. In Part 3, PneuX ETT showed no leak across 8 h for all trials. Overall, leakage volume was lower with PU (P < .01) and PneuX (P < .01) than with PVC cuffs, regardless of shape, and lower with lubrication and/or PEEP (all P < .01). In each HVLP cuff, lubrication alone eliminated leak at an intracuff pressure of ≤30 cm H2O.ConclusionsThe PneuX cuff generally exerted acceptable tracheal wall pressure, but the tracheal wall pressure monitor allowed pressures exceeding 30 cm H2O in some trials and was the only ETT to prevent leak in all tests. For HVLP cuffs, leak was reduced by PU and PEEP and eliminated by lubrication.Copyright © 2017 by Daedalus Enterprises.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…