• Respiratory care · Feb 2017

    FIO2 in an Adult Model Simulating High-Flow Nasal Cannula Therapy.

    • Yusuke Chikata, Mutsuo Onodera, Jun Oto, and Masaji Nishimura.
    • Medical Equipment Center, Tokushima University Hospital, 2-50-1 Kuramoto, Tokushima 770-8503, Japan.
    • Respir Care. 2017 Feb 1; 62 (2): 193-198.

    BackgroundHigh-flow nasal cannula therapy (HFNC) is widely used for patients with acute respiratory failure. HFNC has a number of physiological effects. Although FIO2 is considered to be constant, because HFNC is an open system, FIO2 varies according to inspiratory flow, tidal volume (VT), and HFNC gas flow. We investigated the influence of HFNC gas flow and other respiratory parameters on FIO2 during HFNC.MethodsWe evaluated an HFNC system and, for comparison, a conventional oxygen therapy system. The HFNC apparatus was composed of an air/oxygen blender, a heated humidifier, an inspiratory limb, and small, medium, and large nasal prongs. HFNC gas flow was set at 20, 40, and 60 L/min, and FIO2 was set at 0.3, 0.5, and 0.7. We measured FIO2 for 1-min intervals using an oxygen analyzer and extracted data for the final 3 breaths of each interval. Spontaneous breathing was simulated using a mechanical ventilator connected to the muscle compartment of a model lung. The lung compartment passively moved with the muscle compartment, thus inspiring ambient air via a ventilator limb. With a decelerating flow waveform, simulated VT was set at 300, 500, and 700 mL, breathing frequency at 10 and 20 breaths/min, and inspiratory time at 1.0 s.ResultsWith HFNC gas flow of 20 and 40 L/min, at all set FIO2 values, inspiratory oxygen concentration varied with VT (P < .001). As the set value for FIO2 increased, the difference between set FIO2 and measured FIO2 increased. Neither breathing frequency nor prong size influenced FIO2 .ConclusionsDuring HFNC with simulated spontaneous breathing, when HFNC gas flow was 60 L/min, measured FIO2 was similar to set FIO2 at 0.3 and 0.5, whereas at 0.7, as VT increased, measured FIO2 decreased slightly. However, at 20 or 40 L/min, changes in VT related with deviation from set FIO2 .Copyright © 2017 by Daedalus Enterprises.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…