• Respiratory care · Dec 2016

    Reliability of Displayed Tidal Volume in Healthy and Surfactant-Depleted Piglets.

    • A Cecilia Mendiondo Luedloff, Tracy L Thurman, Shirley J Holt, Shasha Bai, Mark J Heulitt, and Sherry E Courtney.
    • Neonatology Section.
    • Respir Care. 2016 Dec 1; 61 (12): 1605-1612.

    BackgroundVolutrauma has been established as the key factor in ventilator-induced lung injury and can only be avoided if tidal volume (VT) is accurately displayed and delivered. The purpose of this study was to investigate the accuracy of displayed exhaled VT in a ventilator commonly used in small infants with or without a proximal flow sensor and using 3 methods to achieve a target VT in both a healthy and lung-injured neonatal pig model.MethodsThis was a prospective animal study utilizing 8 male pigs, approximately 2.0 kg (range 1.8-2.2 kg). Intubated, sedated, neonatal pigs were studied with both healthy and injured lungs using the Servo-i ventilator. In pressure-regulated volume control, both with and without a proximal flow sensor, we used 3 methods to set VT: (1) circuit compliance compensation (CCC) on, set VT 6-8 mL/kg; (2) CCC off, calculated VT using the manufacturer's circuit compliance factor; and (3) CCC off, set VT 10-12 mL/kg to approximate a target VT of 6-8 mL/kg. Ventilator-displayed exhaled VT measurements were compared with exhaled VT measured at the airway opening by a calibrated pneumotachograph. Bland-Altman plots were constructed to show the level of agreement between the two.ResultsCCC improved accuracy and precision of displayed exhaled VT when the sensor was not used, more markedly in the lung-injured model. Without CCC, the sensor improved accuracy and precision of displayed exhaled VT, again more markedly in the lung-injured model.ConclusionsWhen the Servo-i ventilator is used in neonates, CCC or the in-line sensor should be employed due to the large positive bias and imprecision seen with CCC off and no sensor in-line.Copyright © 2016 by Daedalus Enterprises.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.