    Is it "in" or "out"? The Optimal Fluoroscopic Views for Intraoperative Determination of Proper Lateral Mass Screw Placement.

    • Sang Bum Kim, John M Rhee, Byung Hak Oh, You Gun Won, Yousun Jung, Kun Young Park, William C Hutton, and Chulmin Kim.
    • *Department of Orthopaedic Surgery, Konyang University Hospital, Daejeon, Korea †Department of Orthopaedic Surgery, Emory Spine Center, Atlanta, GA ‡Department of Orthopaedic Surgery, Daejeon Veterans Hosp... more ital, Daejeon, Korea §Department of Mathematics, University of West Georgia, Carrollton, GA. less
    • Spine. 2017 Jul 15; 42 (14): 1039-1043.

    Study DesignCadaveric.ObjectiveDetermine optimal fluoroscopic views for detecting cervical lateral mass screw (LMS) violations.Summary Of Background DataSingle plane intraoperative x-rays are commonly used but frequently inadequate due to its complex trajectory. Fluoroscopy can be taken in multiple planes, but the ideal fluoroscopic view to assess malposition is not known: depending on the view, any given screw may look "in" or "out."MethodsC3-6 LMS were inserted in three cadavers. To evaluate neuroforaminal violation, LMS were inserted into the foramen with the tip penetrating the anterior cortex by 0, 2, and 4 mm. To assess facet joint violation, LMS were inserted toward the subjacent facet joint with the tip penetrating the anterior cortex by 0 and 2 mm. Fluoroscopic views were taken 0°, 10°, 20°, 30°, and 40° to the lateral plane. Views were independently evaluated by three blinded spine surgeons.ResultsTwenty-degree oblique view correctly identified a 2 mm penetration into the neuroforamen in 79%, and a 4 mm penetration in 86%, for a sensitivity of 83% and specificity of 90%. Thirty-degree view had lower sensitivity (76%) but slightly higher specificity (93%). Twenty-degree and 30° views were significantly more sensitive than the other views. Zero-degree view correctly identified a 2 mm penetration into the facet joint in 93%, for a sensitivity of 93% and specificity of 92%. Ten-degree view had lower sensitivity (72%) but higher specificity (100%). The 0° view was significantly more sensitive than the other views.ConclusionTwenty-degree and 30° oblique views significantly provided the most sensitive assessment of LMS potentially violating the neuroforamen, whereas the 0° neutral lateral view significantly provided the most sensitive assessment of facet violations. The specificities were also high (in the 90% range) for these views. We recommend the use of these views intraoperatively when assessing proper placement of LMS fluoroscopically.Level Of EvidenceN/A.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.