J Clin Monit Comput · Dec 2017
Continuous blood pressure monitoring via non-invasive radial artery applanation tonometry and invasive arterial catheter demonstrates good agreement in patients undergoing colon carcinoma surgery.
- Jing Sun, Hanjian Chen, Jun Zheng, Bin Mao, Shengmei Zhu, and Jingyi Feng.
- Department of Clinical Engineering, The First Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, People's Republic of China.
- J Clin Monit Comput. 2017 Dec 1; 31 (6): 1189-1195.
PurposeRadial artery applanation tonometry (RAAT) has been developed and utilized for continuous arterial pressure monitoring. However, evidence is lacking to clinically verify the RAAT technology and identify appropriate patient groups before routine clinical use. This study aims to evaluate the RAAT technology by comparing systolic blood pressure (SBP), mean blood pressure (MBP) and diastolic blood pressure (DBP) values in patients undergoing colon carcinoma surgery.MethodsBlood Pressure (BP) values obtained via RAAT (TL-300, Tensys Medical Inc., San Diego, CA, USA) and conventional arterial catheterization from 30 colon carcinoma surgical patients were collected and compared via Bland-Atman method, linear regression and 4-quadrant plot concordance analysis.ResultsFor SBPs, MBPs and DBPs, means of the differences (±standard deviation; 95% limits of agreement) were -0.9 (±7.6; -15.7 to 13.9) mmHg, 3.1 (±6.5; -9.6 to 15.8) mmHg and 4.3 (±7.4; -10.3 to 18.8) mmHg, respectively. Linear regression coefficients of determination were 0.8706 for SBPs, 0.8353 for MBPs and 0.6858 for DBPs. Four-quadrant concordance correlation coefficients were 0.8740, 0.8522 and 0.7108 for SBPs, MBPs and DBPs, respectively.ConclusionsA highly selected patient collective undergoing colon carcinoma surgery was studied. BP measurements obtained via the TL-300 had clinically acceptable agreement with that acquired invasively using an arterial catheter. For use in clinical routine, it is necessary to take measures for improvement regarding movement artifacts and dilution of noise. A large sample size of patients under various conditions is also needed to further evaluate the RAAT technology before clinically routine use.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.