• Anesthesia and analgesia · Jul 2017

    Multicenter Study

    Bispectral Index Can Reliably Detect Deep Sedation in Mechanically Ventilated Patients: A Prospective Multicenter Validation Study.

    • Zhu-Heng Wang, Han Chen, Yan-Lin Yang, Zhong-Hua Shi, Qing-Hua Guo, Yu-Wei Li, Li-Ping Sun, Wei Qiao, Guan-Hua Zhou, Rong-Guo Yu, Kai Yin, Xuan He, Ming Xu, Laurent J Brochard, and Jian-Xin Zhou.
    • From the *Department of Critical Care Medicine, Beijing Tiantan Hospital, and †Department of Critical Care Medicine, Daxing Teaching Hospital, Capital Medical University, Beijing, China; ‡Surgical Intensive Care Unit, Fujian Provincial Clinical College Hospital, Fujian Medical University, Fuzhou, Fujian, China; §Intensive Care Unit, Beijing Electric Power Hospital, Capital Medical University, Beijing, China; ‖Keenan Research Centre, St Michael's Hospital, Toronto, Canada; and ¶Interdepartmental Division of Critical Care Medicine, University of Toronto, Toronto, Canada.
    • Anesth. Analg. 2017 Jul 1; 125 (1): 176-183.

    BackgroundExcessively deep sedation is prevalent in mechanically ventilated patients and often considered suboptimal. We hypothesized that the bispectral index (BIS), a quantified electroencephalogram instrument, would accurately detect deep levels of sedation.MethodsWe prospectively enrolled 90 critically ill mechanically ventilated patients who were receiving sedation. The BIS was monitored for 24 hours and compared with the Richmond Agitation Sedation Scale (RASS) evaluated every 4 hours. Deep sedation was defined as a RASS of -3 to -5. Threshold values of baseline BIS (the lowest value before RASS assessment) and stimulated BIS (the highest value after standardized assessment) for detecting deep sedation were determined in a training set (45 patients, 262 RASS assessments). Diagnostic accuracy was then analyzed in a validation set (45 patients, 264 RASS assessments).ResultsDeep sedation was only prescribed in 6 (6.7%) patients, but 76 patients (84.4%) had at least 1 episode of deep sedation. Thresholds for detecting deep sedation of 50 for baseline and 80 for stimulated BIS were identified, with respective areas under the receiver-operating characteristic curve of 0.771 (95% confidence interval, 0.714-0.828) and 0.805 (0.752-0.857). The sensitivity and specificity of baseline BIS were 94.0% and 66.5% and of stimulated BIS were 91.0% and 66.5%. When baseline and stimulated BIS were combined, the sensitivity, specificity, and clinical utility index were 85.0% (76.1%-91.1%), 85.9% (79.5%-90.7%), and 66.9% (57.8%-76.0%), respectively.ConclusionsCombining baseline and stimulated BIS may help detect deep sedation in mechanically ventilated patients.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.