• Surgical endoscopy · May 2012

    Mesh fixation with a barbed anchor suture results in significantly less strangulation of the abdominal wall.

    • Calvin Lyons, Rohan Joseph, Nilson Salas, Patrick R Reardon, Barbara L Bass, and Brian J Dunkin.
    • Department of Surgery, Methodist Institute for Technology Innovation and Education (MITIE), The Methodist Hospital, 6670 Bertner Ave, Houston, TX 77030, USA. Clyons@tmhs.org
    • Surg Endosc. 2012 May 1; 26 (5): 1254-7.

    BackgroundLaparoscopic ventral hernia repair using an underlay mesh frequently requires suture fixation across the abdominal wall, which results in significant postoperative pain. This study investigates the utility of a novel mesh fixation technique to reduce the strangulation force on the abdominal wall.MethodsMultiple 2-cm(2) pieces of polyester mesh (Parietex Composite, Covidien) were placed as an underlay against a porcine abdominal wall. Fixation was accomplished using either the standard 0-polyglyconate or the 0-polyglyconate barbed anchor suture designed to hold in tissue without the need to tie a knot (V-Loc 180; Covidien). Suture fixation began with a stab wound incision in the skin. A suture-passing device then was used to pass the suture across the abdominal wall and through the mesh. The suture passer was removed and reintroduced through the same stab wound incision but at a different fascial entry point 1.5 cm away. The tail of the suture was grasped and pulled up through both the mesh and the abdominal wall, creating a full-thickness U-stitch. One tail of the suture was attached to a tensiometer, and the strangulation force on the abdominal wall was measured while the suture was tied (standard) or looped (barbed). To compare pullout force, the tensiometer was attached to either the mesh or the suture, and traction was applied until material failure or suture pull through. Results are expressed as mean ± standard deviation. Comparisons were performed using Student's t-test.ResultsEight pieces of mesh were placed for each suture. The average force required to secure the barbed suture (0.59 ± 0.08 kg) was significantly less than the force needed to secure the standard suture (2.17 ± 0.58 kg) (P < 0.0001). Table 1 compares the suture pullout forces with the mesh failure forces. Although the pullout force for the standard suture is significantly greater than for the barbed suture, both sutures have a pullout strength significantly greater than the mesh failure force. Table 1 Suture fixation forces for standard and barbed sutures Suture fixation force (kg) Standard suture 2.17 ± 0.58 Barbed suture 0.59 ± 0.08 P < 0.0001ConclusionsA barbed anchor suture used to secure mesh to the abdominal wall requires nearly 75% less strangulation force than a standard monofilament suture while still providing significantly greater pullout force than that required for the mesh to tear and fail. This method of mesh fixation should result in less postoperative pain and warrants a clinical trial.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…