• Arch. Pathol. Lab. Med. · Jul 2014

    Reliability of point-of-care capillary blood glucose measurements in the critical value range.

    • Ron B Schifman, Tan T Nguyen, and Susan T Page.
    • From the Department of Pathology and Laboratory Medicine, Southern Arizona VA Healthcare System (Drs Schifman, Nguyen, Ms Page), and the Department of Pathology, University of Arizona (Drs Schifman, Nguyen).
    • Arch. Pathol. Lab. Med. 2014 Jul 1; 138 (7): 962-6.

    ContextPoint-of-care glucose (POCG) testing on capillary blood specimens is central to maintaining glycemic control in patients with diabetes. Although there are known performance issues with POCG methods, especially for maintaining tight glucose control, there is little information about the accuracy of results in the critical ranges that may involve life-threatening conditions.ObjectivesTo evaluate the reliability of POCG measurements in critical, high (>600 mg/dL) and low (<40 mg/dL) ranges.DesignOne-year retrospective analysis of POCG (ACCU-CHEK glucose meter, Roche Diagnostics Corporation, Indianapolis, Indiana) results for routine patient care were obtained. The frequency and accuracy of repeat testing after critical POCG results was analyzed. A convenience sample of noncritical capillary POCG measurements retested on venous blood specimens by another point-of-care device (RAPIDPoint 405 analyzer, Siemens Medical Solutions USA, Malvern, Pennsylvania) was also evaluated.ResultsCritical values were observed in 2.4 per 1000 POCG tests (256 of 105,928; 0.24%), with the highest rate (76 of 2289; 3.32%) from the emergency department. Twice as many critical high values as critical low values were seen. Nearly 80% of critical POCG tests (204 of 256) were repeated within 10 minutes. Of these 204 repeat measurements, 112 (54.9%) met accuracy criteria (±15 mg/dL of low and ±20% of high initial values). Accuracy was significantly higher when retesting was performed on the same meter or by the same operator (P ≤ .05). Comparison of capillary and venous POCG testing of noncritical results showed no significant difference (P = .95), with 89.8% (125 of 139) meeting accuracy criteria.ConclusionsPOCG measurements in the critical range are frequently erroneous, which is likely caused by preanalytic factors associated with sampling capillary blood. POCG testing practices should include retesting to confirm critical results.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…