• Pediatric emergency care · Dec 2021

    Case Reports

    Inappropriate Shocks and Power Delivery Using Adult Automatic External Defibrillator Pads in a Pediatric Patient.

    • Shai Tejman-Yarden, Uriel Katz, Marina Rubinstein, Yehuda Attias, Reem Yahia, David Mishali, and Michael Glikson.
    • From the Edmond J. Safra International Center for Congenital Heart Diseases.
    • Pediatr Emerg Care. 2021 Dec 1; 37 (12): e1708e1710e1708-e1710.

    AbstractThe use of automatic external defibrillators (AEDs) during pulseless resuscitations is considered safe and reliable, and was established as part of the guidelines in out-of-hospital events. Based on extensive studies, the use of the standard AED is now indicated in every age group with a preference of pediatric pad application for small children and babies. If unavailable, adult pads are recommended. We report a case of 2 inappropriate AED shocks that were delivered to a neonate during a pulseless resuscitation after application of adult pads. The 3.6-kg patient received 2 shocks, over 200 J each, for sinus bradycardia that was not detected by the device. Although treated inappropriately with high voltage, no cardiac or skin sequelae were detected, and the patient had normal cardiac and neurological development later on.Copyright © 2017 Wolters Kluwer Health, Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.