• Am. J. Crit. Care · Mar 2017

    Review

    Update on Effectiveness of an Electromagnetic Feeding Tube-Placement Device in Detecting Respiratory Placements.

    • Norma A Metheny and Kathleen L Meert.
    • Norma A. Metheny is a professor of nursing and holds the Dorothy A. Votsmier Endowed Chair in Nursing at Saint Louis University School of Nursing, St Louis, Missouri. Kathleen L. Meert is a professor of medicine in the Department of Pediatrics, Wayne State University, and chief, Division of Critical Care Medicine, Children's Hospital of Michigan, Detroit, Michigan. methenna@slu.edu.
    • Am. J. Crit. Care. 2017 Mar 1; 26 (2): 157-161.

    BackgroundRadiography is the accepted gold standard for testing feeding tube placement; however, an electromagnetic tube-placement device (ETPD) is sometimes used in lieu of radiography for this purpose. High success rates have been reported when the device was used by well-trained individuals. However, authors previously described 20 cases that occurred between 2007 and 2012 in which clinicians voluntarily reported inability to detect inadvertent tube insertions in the respiratory tract while using an ETPD.ObjectiveTo describe case reports to the US Food and Drug Administration's Manufacturer and User Facility Device Experience (MAUDE) database between 2013 and 2015 regarding inadvertent respiratory placement of feeding tubes by operators using an ETPD.MethodsThe MAUDE database was searched for cases dated from January 1, 2013, through December 31, 2015, along with selected brand names. A total of 34 cases (25 after removal of duplicates) were located in which a feeding tube was inserted into the respiratory tract during insertions assisted by an ETPD.ResultsSites of the malpositioned tubes included the right lung (n = 13), left lung (n = 6), unspecified lung (n = 4), and bronchus (n = 2). A pneumothorax occurred in 17 of the 25 misplacements; feedings were administered in 6 cases.ConclusionsMany case reports involved clinicians failing to recognize tube misplacements in the respiratory tract while using an ETPD. These reports provide evidence that not all clinicians can use the device effectively to detect malpositioned tubes. Thus, one must continue to question the wisdom of eliminating radiographic confirmation of tube position before starting feedings.©2017 American Association of Critical-Care Nurses.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…