J. Thorac. Cardiovasc. Surg. · Jun 2017
Comparative StudyEffect of transcatheter aortic valve size and position on valve-in-valve hemodynamics: An in vitro study.
- Ali N Azadani, Michael Reardon, Matheus Simonato, Gabriel Aldea, Georg Nickenig, Ran Kornowski, and Danny Dvir.
- Department of Mechanical and Materials Engineering, University of Denver, Denver, Colo.
- J. Thorac. Cardiovasc. Surg. 2017 Jun 1; 153 (6): 1303-1315.e1.
ObjectiveTranscatheter heart valve implantation in failed aortic bioprostheses (valve-in-valve [ViV]) is an increasingly used therapeutic option for high-risk patients. However, high postprocedural gradients are a significant limitation of aortic ViV. Our objective was to evaluate Medtronic CoreValve Evolut R ViV hemodynamics in relation to the degree of device oversizing and depth of implantation.MethodsEvolut R devices of 23 and 26 mm were implanted within 21-, 23-, and 25-mm Hancock II bioprostheses. Small and gradual changes in implantation depth were attempted. Hemodynamic testing was performed in a pulse duplicator under ISO-5840 standard.ResultsA total of 47 bench-testing experiments were performed. The mean gradient of the 26-mm Evolut R in 23- and 25-mm Hancock II was lower than 23-mm Evolut R (P < .001). However, the mean gradient of 26-mm Evolut R in 21-mm Hancock II bioprostheses R (ranging from 21.30 ± 0.23 to 24.30 ± 0.22 mm Hg) was worse than 23-mm Evolut R (ranging from 15.94 ± 0.18 to 20.35 ± 0.16 mm Hg, P < .001). Furthermore, our results suggest that supra-annular implantation of 23-mm and 26-mm Evolut R devices within the bioprostheses can lead to lower gradient and improved leaflet coaptation. Regardless of implantation depth, superior transvalvular gradient is expected with 26-mm Evolut R than 23-mm Evolut R in a nonstenotic Hancock II with a true internal diameter > 17.5 mm.ConclusionsThe current comprehensive bench-testing assessment demonstrates the importance of both transcatheter heart valve size and device position for the attainment of optimal hemodynamics during ViV procedures. Additional in vitro testing may be required to develop hemodynamics-based guidelines for device sizing in ViV procedures in degenerated surgical bioprostheses.Copyright © 2017 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.