Comparative Study Observational StudyAn In Vitro Assessment of the Efficacy of Various IV Cannulas for the Rapid IV Fluid Administration.
- Mineto Kamata, Hina Walia, Mumin Hakim, Dmitry Tumin, and Joseph D Tobias.
- 1Department of Anesthesia, Saitama Children's Medical Center, Saitama, Japan. 2Department of Anesthesiology & Pain Medicine, Nationwide Children's Hospital, Columbus, OH. 3Department of Anesthesiology & Pain Medicine, The Ohio State University College of Medicine, Columbus, OH. 4Department of Pediatrics, The Ohio State University College of Medicine, Columbus, OH.
- Pediatr Crit Care Me. 2017 May 1; 18 (5): e224-e228.
ObjectivesThe current study prospectively evaluates the administration of fluid through commonly used vascular cannulas of various length and diameter.DesignObservational, in vitro experiment.SettingNot applicable.SubjectsNo human subjects.InterventionsNone.Measurements And Main ResultsFluid (500 mL) was administered via gravity flow and with pressure assistance (pressure bag set at 300 mm Hg) through various vascular cannulas including peripheral IV catheters (22 gauge, 1 inch; 20 gauge, 1.16 and 1.88 inch; and 18 gauge, 1.16 and 1.88 inch), 3-Fr central line lumens (lengths 50, 80, and 120 mm), and a 4-Fr, 10 cm Micropuncture catheter (Cook Medical, Bloomington, IN). During gravity flow, drain time decreased by approximately 50% when moving from a 22 to 20 to 18 gauge cannula and increased by approximately 20% as the catheter length increased from 50 to 80 to 120 cm. Flow rates were highest with the Micropuncture catheter, which achieved a drain time of 2.7 minutes for the 500 mL bag when infused with pressure assistance.ConclusionsIn general, the delivery of crystalloid solutions through commonly used vascular cannulas was fastest with larger diameter and shorter length cannulas. Pressure-assisted flow was able to partially compensate for the increased resistance with smaller and longer catheters. The unique design of the tip of the Micropuncture catheter compensated for the increased length and allowed for rapid fluid delivery. This design may compensate for the increased length when longer catheters are needed for ultrasound-guided placement.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.