• Anesthesia and analgesia · Jul 2017

    Randomized Controlled Trial

    Closed-Loop Feedback Computer-Controlled Phenylephrine for Maintenance of Blood Pressure During Spinal Anesthesia for Cesarean Delivery: A Randomized Trial Comparing Automated Boluses Versus Infusion.

    • Warwick D Ngan Kee, Yuk-Ho Tam, Kim S Khaw, Floria F Ng, and Shara W Y Lee.
    • From the *Department of Anaesthesia and Intensive Care, the Chinese University of Hong Kong, China; and †Department of Health Technology and Informatics, the Hong Kong Polytechnic University, Hung Hom, Hong Kong.
    • Anesth. Analg. 2017 Jul 1; 125 (1): 117-123.

    BackgroundWe previously described the use of closed-loop feedback computer-controlled infusion of phenylephrine for maintaining blood pressure (BP) during spinal anesthesia for cesarean delivery. In this study, we report a modified system in which phenylephrine is delivered by intermittent boluses rather than infusion. We hypothesized that the use of computer-controlled boluses would result in more precise control of BP compared with infusions.MethodsTwo hundred fourteen healthy patients having spinal anesthesia for elective cesarean delivery were randomized to have their systolic BP maintained by phenylephrine administered by computer-controlled continuous infusion or computer-controlled intermittent boluses. From induction of anesthesia until the time of uterine incision, a noninvasive BP monitor was set to cycle at 1-minute intervals. In the infusion group, the infusion rate was automatically adjusted after each BP measurement using a previously described algorithm. In the bolus group, the algorithm was modified so that the mass of drug that would have been delivered over 1 minute was instead injected as a rapid intravenous bolus after each BP measurement. The precision of BP control was assessed using performance error calculations and compared between groups, with the primary outcome defined as median absolute performance error, and the latter being a measure of inaccuracy showing an average of the magnitudes of the differences of measured BP values above or below the target values.ResultsThe precision of BP control was greater, as shown by smaller values for median absolute performance error, in the bolus group (median 4.38 [quartiles 3.22, 6.25] %) versus the infusion group (5.39 [4.12, 7.04] %, P = .008). In the bolus group, phenylephrine consumption was smaller; this was associated with smaller values for median performance error compared with the continuous infusion group (P < .001), which indicates that values for systolic BP, averaged over time, were slightly lower in the bolus group. There were no differences in cardiac output, nausea or vomiting, or neonatal outcome between groups.ConclusionsWe confirmed the hypothesis that BP control was more precise when computer-controlled phenylephrine was delivered using intermittent boluses rather than continuous infusion. However, the difference between groups was small and was not associated with any difference in clinical outcomes. In the infusion group, greater doses of phenylephrine were delivered, which was related to the time taken for the noninvasive BP monitor to complete measurements. The use of intermittent boluses may be a useful alternative in the design of closed-loop vasopressor systems.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…