Arch. Dis. Child. Fetal Neonatal Ed. · Jan 2017
Clinical TrialClinical evaluation of a novel adaptive algorithm for automated control of oxygen therapy in preterm infants on non-invasive respiratory support.
- Gemma K Plottier, Kevin I Wheeler, Sanoj K M Ali, Omid Sadeghi Fathabadi, Rohan Jayakar, Timothy J Gale, and Peter A Dargaville.
- Department of Paediatrics, Royal Hobart Hospital, Hobart, Tasmania, Australia.
- Arch. Dis. Child. Fetal Neonatal Ed. 2017 Jan 1; 102 (1): F37-F43.
ObjectiveTo evaluate the performance of a novel rapidly responsive proportional-integral-derivative (PID) algorithm for automated oxygen control in preterm infants with respiratory insufficiency.DesignInterventional study of a 4-hour period of automated oxygen control compared with combined data from two flanking periods of manual control (4 hours each).SettingNeonatal intensive care unit.ParticipantsPreterm infants (n=20) on non-invasive respiratory support and supplemental oxygen, with oxygen saturation (SpO2) target range 90%-94% (manual control) and 91%-95% (automated control). Median gestation at birth 27.5 weeks (IQR 26-30 weeks), postnatal age 8.0 (1.8-34) days.InterventionAutomated oxygen control using a standalone device, receiving SpO2 input from a standard oximeter and computing alterations to oxygen concentration that were actuated with a modified blender. The PID algorithm was enhanced to avoid iatrogenic hyperoxaemia and adapt to the severity of lung dysfunction.Main Outcome MeasureProportion of time in the SpO2 target range, or above target range when in air.ResultsAutomated oxygen control resulted in more time in the target range or above in air (manual 56 (48-63)% vs automated 81 (76-90)%, p<0.001) and less time at both extremes of oxygenation. Prolonged episodes of hypoxaemia and hyperoxaemia were virtually eliminated. The control algorithm showed benefit in every infant. Manual changes to oxygen therapy were infrequent during automated control (0.24/hour vs 2.3/hour during manual control), and oxygen requirements were unchanged (automated control period 27%, manual 27% and 26%, p>0.05).ConclusionsThe novel PID algorithm was very effective for automated oxygen control in preterm infants, and deserves further investigation.Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.