J Clin Monit Comput · Aug 2018
Randomized Controlled TrialDesflurane usage during anesthesia with and without N2O using FLOW-i Automatic Gas Control with three different wash-in speeds.
- Robrecht De Medts, Rik Carette, Andre M De Wolf, and Hendrickx Jan F A JFA Department of Anesthesiology, Intensive Care and Pain Therapy, OLV Hospital, Moorselbaan 164, 9300, Aalst, Belgium..
- Department of Anesthesiology, Intensive Care and Pain Therapy, OLV Hospital, Moorselbaan 164, 9300, Aalst, Belgium. robrechtdemedts@hotmail.com.
- J Clin Monit Comput. 2018 Aug 1; 32 (4): 763-769.
AbstractAGC® (Automatic Gas Control) is the FLOW-i's automated low flow tool (Maquet, Solna, Sweden) that target controls the inspired O2 (FIO2) and end-expired desflurane concentration (FAdes) while (by design) exponentially decreasing fresh gas flow (FGF) during wash-in to a maintenance default FGF of 300 mL min-1. It also offers a choice of wash-in speeds for the inhaled agents. We examined AGC performance and hypothesized that the use of lower wash-in speeds and N2O both reduce desflurane usage (Vdes). After obtaining IRB approval and patient consent, 78 ASA I-II patients undergoing abdominal surgery were randomly assigned to 1 of 6 groups (n = 13 each), depending on carrier gas (O2/air or O2/N2O) and wash-in speed (AGC speed 2, 4, or 6) of desflurane, resulting in groups air/2, air/4, air/6, N2O/2, N2O/4, and N2O/6. The target for FIO2 was set at 35%, while the FAdes target was selected so that the AGC displayed 1.3 MAC (corrected for the additive affect of N2O if used). AGC was activated upon starting mechanical ventilation. Varvel's criteria were used to describe performance of achieving the targets. Patient demographics, end-expired N2O concentration, MAC, FGF, and Vdes were compared using ANOVA. Data are presented as mean ± standard deviation, except for Varvel's criteria (median ± quartiles). Patient demographics did not differ among the groups. Median performance error was -2-0% for FIO2 and -3-1% for FAdes; median absolute performance error was 1-2% for FIO2 and 0-3% for FAdes. MAC increased faster in N2O groups, but total MAC decreased 0.1-0.25 MAC below that in the O2/air groups after 60 min. The effect of wash-in speed on Vdes faded over time. N2O decreased Vdes by 62%. AGC performance for O2 and desflurane targeting is excellent. After 1 h, the wash-in speeds tested are unlikely to affect desflurane usage. N2O usage decreases Vdes proportionally with its reduction in FAtdes.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.