- Christof Karmonik, Jeff R Anderson, Saba Elias, Richard Klucznik, Orlando Diaz, Yi Jonathan Zhang, Robert G Grossman, and Gavin W Britz.
- Cerebrovascular Center, Department of Neurosurgery, Houston Methodist Hospital, Houston, Texas, USA; MRI Core, Houston Methodist Research Institute, Houston, Texas, USA. Electronic address: ckarmonik@houstonmethodist.org.
- World Neurosurg. 2017 Sep 1; 105: 775-782.
ObjectiveHemodynamics in cerebral aneurysms are currently investigated toward clinical efficacy using nonstandardized computational simulation techniques. At the same time, flow patterns and velocities are accessible by 4-dimensional phase contrast magnetic resonance imaging (4D pcMRI). Complexity of protocol design and imaging duration has limited the use of this technique in clinical imaging. A new approach is presented to overcome these limitations.MethodsThree-dimensional (3D) replicas of 2 cerebral aneurysms were fabricated by fused deposition prototyping (3D printing) and imaged using 4D pcMRI while connected to a magnetic resonance imaging-compatible continuous flow loop. Acquisition parameters were optimized with imaging times not to exceed 10 minutes. Six patients harboring cerebral aneurysms with sizes ranging from 4.7 to 13.8 mm were imaged with the optimized 4D pcMRI protocol. After treatment with the pipeline embolization device (PED), 4D pcMRI examinations were repeated in 3 patients.ResultsIn all cases, major flow patterns were visualized well; smaller aneurysms posed a challenge because of limited spatial resolution, whereas larger aneurysms contained regions of low velocity resulting in limited contrast in the flow-sensitive images. After PED placement, ordered aneurysmal flow was disrupted and intra-aneurysmal velocity was reduced on average by 24.5% (range, 12.9-31.5%). Exploratory statistical analysis yielded a positive significant correlation (P < 0.01) between changes in inflow velocity and posttreatment intra-aneurysmal flow velocity.Conclusions4D pcMRI flow imaging in cerebral aneurysms within a time frame suitable for clinical imaging applications is feasible with optimized acquisition parameters, thereby enabling quantification of intra-aneurysmal flow changes after flow diverter device treatment.Copyright © 2017 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.