    Investigation of Two Prototypes of Novel Noncontact Technologies for Automated Real-Time Capture of Incremental Drug Administration Data From Syringes.

    • Benjamin Eagle, David J Williams, and John Dingley.
    • From the *Morriston Hospital, Department of Anaesthetics, ABM University Health Board, Swansea, United Kingdom; †Welsh Centre for Burns, ABM University Health Board, Swansea, United Kingdom; and ‡Swansea University College of Medicine, Swansea, United Kingdom.
    • Anesth. Analg. 2017 Aug 1; 125 (2): 458-466.

    BackgroundAn ideal electronic anesthesia recording system would be capable of not only recording physiological data but also injectable drug doses given, including those given incrementally from one syringe, without recourse to manual data entry. We compared 2 prototype devices which wirelessly recognized individual syringes and measured changes in their plunger positions via 2 different optical noncontact means, allowing calculation of incremental drug doses given.MethodsBoth devices incorporated a radio-frequency identification reader, which wirelessly read a unique code from a radio-frequency identification tag within syringe drug labels. A custom-designed cradle oriented any inserted 1-mL to 20-mL syringe in a repeatable position. The "laser" device had a moving laser beam broken by the end of the syringe plunger. The infrared (IR) device measured time of travel of IR light from a sender to a syringe plunger and back to a receiver. Both devices could therefore determine the drug and volume administered since the previous occasion when any syringe had been used. For each syringe size of 1, 2, 5, 10, and 20 mL, 121 plunger-length measurements were made over their full range, with each machine against a reference method of water filling and weighing using a randomized de Bruijn sequence.ResultsFor every syringe size, the laser device showed greater accuracy and precision, lower bias, and narrower limits of agreement (95% confidence intervals = bias ± 1.96 SD) than the IR device when compared to the reference method. For all syringe sizes, the range of bias was -0.05 to 0.32 mL for the laser and -2.42 to 1.38 mL for the IR. Lin concordance correlation coefficient values for the IR versus reference methods ranged from 0.6259 to 0.9255, with the lowest coefficients seen in syringes with the shortest distance of plunger travel (2 and 5 mL), while in laser versus reference comparisons, these coefficients were similar (0.9641-0.9981) over all syringe lengths.ConclusionsBoth devices measured syringe volume changes, demonstrating potential for measuring incremental drug doses, recording these, and also the time of each measurement. The IR device had no moving parts, which would be advantageous in a clinical situation. However, the current embodiment was not deemed accurate enough for clinical use, potentially remediable through improvements in hardware and software design. The laser device showed high accuracy and precision over all syringe sizes and contained volumes, and was considered potentially accurate enough for clinical use with suitable development.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.