• J. Appl. Physiol. · Dec 1994

    Evaluation of a symmetrically disposed Pitot tube flowmeter for measuring gas flow during exercise.

    • J Porszasz, T J Barstow, and K Wasserman.
    • Department of Medicine, University of California, Los Angeles Harbor Medical Center, Torrance 90509.
    • J. Appl. Physiol. 1994 Dec 1; 77 (6): 2659-65.

    AbstractWe evaluated the effect of airflow and gas composition on the linearity of measurement of airflow by a new disposable flowmeter. The flowmeter is based on the principle of differential pressure measurement across two symmetrically disposed Pitot tubes. Nonlinearities arising from the pressure-to-airflow relationship and sensitivity to changes in gas density were linearized with appropriate software and monitoring of the gas composition. With room air used as the respired gas, the measured tidal volume from a piston pump assembly was consistently within 1-2% of the target tidal volume for each of five flowmeters tested across physiological ranges of flow. Changing gas densities by varying concentrations of O2, CO2, and N2 led to errors in tidal volume measurement that ranged up to 6-8%. However, because the errors were predictable, they were corrected by software to within 0.6% of the target volume. Measurement of minute ventilation during exercise was within 1-2% of that determined from bag collections. We conclude that this type of flowmeter can accurately measure exercise minute ventilation and has advantages over some other flowmeters because of its ruggedness, reproducibility, and ease of sterilization or replacement compared with other flowmeters.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…