Review Comparative Study
[Biomechanical study of four palmar locking plates and one non-locking palmar plate for distal radius fractures: stiffness and load to failure tests in a cadaver model].
- Lothar Rudig, Isabella Mehling, Daniela Klitscher, Dorothea Mehler, Karl-Josef Prommersberger, Pol M Rommens, and Lars Peter Müller.
- Klinik für Unfall-, Hand- und Wiederherstellungschirurgie GP Rüsselsheim, Deutschland. rudig@GP-Ruesselsheim.de
- Biomed Tech (Berl). 2009 Jun 1; 54 (3): 150-8.
AbstractFive different palmar fixation plate designs were compared in a distal radial osteotomy cadaver model with regard to their biomechanical properties. A metaphyseal osteotomy gap of 1 cm was performed and the osteosynthesis was plated according to the manufacturer's instructions. Axial load was applied to the construct by a pneumatic material testing machine. Five implant groups with eight cadavers each were tested concerning stiffness. None of the constructs developed deformity and movement of the fracture gap larger than 2 mm with a load of 100 N. Increasing the load to 250 N revealed significant differences in stiffness and failure load between the different plates. The mean stiffness under axial load (mean+/-standard deviation) was 356.4+/- 138.6 N/mm for the radius correction plate without lateral tongue, 299.7+/-86.3 N/mm for the radius correction plate with lateral tongue, 132.8+/-41.5 N/mm for the distal volar radius plate, 112.5+/-40.2 N/mm for the 3.5 mm titanium locking compression plate and 91.9+/-29.2 N/mm for the standard stainless steel 3.5 mm T-Plate. The non-angular stable implant (STP plate) had the lowest stiffness. Unexpectedly, there were differences over 100% concerning the stiffness between the at first glance nearly similar angular stable implants. Additionally, a review of the literature concerning biomechanical investigations of the distal radial fracture was performed.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.